Table 12. Meter Accuracy (10 V to 1000 V)
Equipment Setup

Calibration Procedure For Model 230B, 860A, 865A, 290A, B, 801, And Sp2979 Page Navigation
  23    24    25    26    27  28  29    30    31    32    33  
img
. . . . . . .
TB 9-4931-217-40
Figure 5. Dc generator detector - rear interior view - adjustment locations.
29. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
SECTION VII
CALIBRATION PROCESS FOR AC/DC GENERATOR DETECTOR, MODEL 865A
30. Preliminary Instructions
a. The instructions outlined in paragraphs 30 and 31 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance check where applicable.