TB 9-4931-217-40
(2) Connect fixed resistor directly across TI GENERATOR AC OUTPUT terminals
1 and 2.
(3) Connect TI GENERATOR AC OUTPUT terminal 2 to TI DETECTOR AC
INPUT terminal 2.
(4) Press ON/OFF pushbutton to ON. Adjust R14 (fig. 6) until meter indicates null
(low point, not zero).
(5) Set GENERATOR polarity switch alternately between + (positive) and -
(negative). If TI meter indication is not within one minor division in either position,
perform b below.
b. Adjustments (fig. 6). Set GENERATOR polarity switch alternately between +
(positive) and - (negative) while adjusting R14 for same TI meter indication in either
position.
a. Performance Check
(1) Connect function/arbitrary generator 50-ohm output terminal
to
TI
DETECTOR AC INPUT terminals 1 and 2, using cable and termination.
(2) Adjust function/arbitrary generator controls for a 100 V output with frequency
at 1 kHz.
(3) Set DETECTOR SENSITIVITY HIGH/LOW switch to HIGH and rotate
DETECTOR SENSITIVITY control fully cw. If TI meter does not indicate full scale,
perform b below.
(4) Set DETECTOR SENSITIVITY HIGH/LOW switch to LOW.
(5) Adjust function/arbitrary generator controls for a 10 mV output. TI meter will
indicate full scale or near full scale.
b. Adjustments. Adjust R15 GAIN (fig. 6) for a full-scale indication on TI meter.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
36. Preliminary Instructions
a. The instructions outlined in paragraphs 36 and 37 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning