Equipment Setup
100 Ohm Decade Linearization

Calibration Procedure For Model 230B, 860A, 865A, 290A, B, 801, And Sp2979 Page Navigation
  28    29    30    31    32  33  34    35    36    37    38  
TB 9-4931-217-40
(1) GENERATOR OUTPUT +/OFF/- switch to OFF.
(2) GENERATOR POWER LIMIT control to 100 MILLIWATTS.
(3) GENERATOR RANGE, switch to 10 k .
(4) DETECTOR RANGE switch to 30 MICROVOLTS.
(5) MULTIPLIER switch to 1X STANDARD.
(6) DEVIATION RANGE switch to +0.1 ppm.
(7) DEVIATION dial to X0 (verify that index is aligned with 0 (zero).
(8) Resistance decade dials to 009999.
e. Adjust resistance standard no. 1 to 10,000.00 ohms.
38. Deviation Calibration
a. Performance Check
(1) Position controls as listed in (a) through (c) below:
(a) DETECTOR ZERO control for a null indication on TI meter.
(b) DETECTOR RANGE switch to 30 MILLIVOLTS.
GENERATOR OUTPUT +/OFF/- switch to + (positive).
(c)
(2) Adjust resistance decade dials for a null indication on TI meter while decreasing
DETECTOR RANGE switch to 30 MICROVOLTS.
(3) Position controls as fisted in (a) through (d) below:
(a) DETECTOR RANGE switch to 30 MILLIVOLTS.
(b) GENERATOR OUTPUT +/OFF/- switch to OFF.
(c) FUNCTION switch to LEAD ADJ.
(d) GENERATOR OUTPUT +/OFF/- switch to + (positive).
(4) Adjust LEAD ADJ control for a null indication on TI meter while decreasing
DETECTOR RANGE switch to 30 MICROVOLTS.
(5) Position controls as listed in (a) through (d) below:
(a) DETECTOR RANGE switch to 30 MILLIVOLTS.
(b) GENERATOR OUTPUT +/OFF/- switch to OFF.
(c) FUNCTION switch to YOKE ADJ.
(d) GENERATOR OUTPUT +/OFF/- switch to + (positive).
(6) Adjust YOKE ADJ control for a null indication on TI meter while decreasing
DETECTOR RANGE switch to 30 MICROVOLTS.
(7) Position controls as fisted in (a) through (d) below:
(a) DETECTOR RANGE switch to 30 MILLIVOLTS.
(b) GENERATOR OUTPUT +/OFF/- switch to OFF.
33