Equipment Setup
Ac Detector Sensitivity

Calibration Procedure For Model 230B, 860A, 865A, 290A, B, 801, And Sp2979 Page Navigation
  25    26    27    28    29  30  31    32    33    34    35  
img
. . . . .
TB 9-4931-217-40
Figure 6. Model 865A - rear internal view.
32. Ac Generator
a. Performance Check
(1) Set FREQUENCY switch to 1 kHz and rotate GENERATOR POWER LIMIT
control fully cw.
(2) Connect multimeter and frequency counter to TI GENERATOR AC OUTPUT
terminals 1 and 2.
(3) Set GENERATOR ON/OFF switch to ON. If multimeter does not indicate at
least 6 V ac, perform b (1) below. If frequency counter does not indicate between 990 and
1,010 Hz (hertz), perform b (2) below.
b. Adjustments (fig. 6)
(1) Adjust R6 for a 6 V ac indication on multimeter (R).
(2) Adjust C6 for a 1 kHz indication on frequency counter (R).
33. Ac Balance
a. Performance Check
(1) Press ON/OFF pushbutton to OFF and adjust DETECTOR SENSITIVITY
control to midrange.
30