TB 9-4931-383-50
Table 1. Calibration Description
Test Instrument Parameters
Performance Specifications1
Voltage range
10-v range; 0 to 11.111110 v (1-v steps)
100-v range; 0 to 111.11110 v (10-v steps)
1000-v range; 0 to 1111.1110 v (100-v steps)
Resolution Accuracy
0.1 ppm of range (1 v max)
10-v range; (0.003% of setting +10 v)
100-v range; (0.003% of setting +20 v)
1000-v range; (0.003% of setting +40 v)
Over-voltage protection
Automatically disables output voltage if level
exceeds selected value (10 to 110% of selected
range).
Output current
0 to 50 ma (any voltage setting)
Over-current protection
Automatically limits output current to present
level; between 1 and 60 ma.
Line and load regulation
0.0005% of setting or 25 v for 10% change in
line voltage from nominal.
1The
specifications for models 332B and 332B/AF have been derated to the same as model 332A
which are, reflected in this calibration description.
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2
identifies the
specific equipment used in
this
calibration procedure. This equipment is issued with secondary reference standards
calibration set NSN 4931-00-621-7878, and is to be used in performing this procedure.
Alternate items may be used by the calibrating activity when the equipment listed in table
2 is not
available. The
items selected must be
verified to
perform satisfactorily prior to
use
and must bear evidence of current calibration. The equipment must meet or exceed the
minimum use specifications listed in table 2. When accuracies listed in table 2 provide a
four-to-one accuracy ratio between the standard and TI. Where the four-to-one ratio cannot
be met, the accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The
accessories listed in
table 3 are issued as
indicated in
paragraph 4 above and
are to be
used in this
calibration procedure. When necessary, these
items may be substituted by equivalent items unless specifically prohibited.
Table 2. Minimum Specifications of Equipment Required
Manufacturer, Model, and
Item
Common Name
Minimum Use Specifications
Part Number
A1
Range: 105 to 125 vac
General Radio, Model
Accuracy: 1%
WI0MT3AS3 (7910809)
A2
DC TRANSFER
Range: 1 and 10 vdc
John Fluke, Model 730A/AB (MIS-
STANDARD
10358, type 2)
Accuracy: 0.0005% when
transferred from saturated
cells
A3
DECADE RESISTOR
Range: 20 to 200 ohms
Biddle-Gray, Model 601147-1
Accuracy: l
(7910328)
A4
NULL DETECTOR
Range: 0 to 1000 vdc
John Fluke, Model 845AB
Accuracy: N/A
(8205517)
3