TB 9-4931-383-50
NOTE
Unless otherwise specified, all controls and control settings
refer to the TI.
c.
When indications specified in
paragraph 10 are not within
tolerance, perform the
power supply check prior to making adjustments. After adjustments are made, repeat
paragraph 10. Do not
perform power supply check if all
other parameters are within
tolerance.
d. Position TI controls as listed in (1) through (5) below:
(1) VOLTAGE RANGE switch to 10.
(2) Voltage dials to 1.000000.
(3) METER switch to VOLTAGE.
(4) VOLTAGE TRIP control fully clockwise.
(5) CURRENT LIMIT control fully clockwise.
e.
Connect TI to
autotransformer (A1) and
leave connected throughout procedure.
NOTE
The sensitivity switch (B1) has been included in equipment
setup for safety and to allow reversal of connections to null
detector (A4). Detailed operation of the switch is not included
in step-by-step procedure.
f.
Connect autotransformer to a
115-volt ac
source and
adjust for a
115-volt output.
g. Turn POWER switch to STANDBY/RESET and allow at least 1 hour for warm-up.
h. Connect operating voltage source (A5) to voltage regulator (A8), set its POWER
switch to STANDBY/ RESET, and allow to warm-up until called for.
i. It is necessary to know actual value of 1.018 +∆E output of dc transfer standard (A2)
to complete this calibration. If calibration of dc transfer standard is no more than 10 days
old, the value recorded at time of calibration is sufficient. If calibration is more than 10
days old, determine the value by performing (1) through (11) below:
5