TB 9-4931-488-35
Table 5. Output Meter Accuracy (Standard, Options 001 and H66)
Measuring receiver indications
Test instrument
(dBm)
OUTPUT LEVEL
Meter
switch settings
indications
(dBm)
(dB)
Min
Max
+20
-1
+17.5
+20.5
+20
-7
+11.5
+14.5
+20
-10
+8.5
+11.5
+10
+3
+11.5
+14.5
+10
0
+8.5
+11.5
0
+3
+1.5
+4.5
0
0
-1.5
+1.5
Table 6. Output Meter Accuracy (Option 002)
Test
Meter
instrument
indications
Measuring receiver indications
OUTPUT
(dB)
(dBm)
LEVEL
switch
25 MHz
512 MHz
1000 MHz
settings
(dBm)
Min
Max
Min
Max
Min
Max
+20
-2
+16.5
+19.5
+16.0
+20.0
---
---
+20
-7
+11.5
+14.5
+11.0
+15.0
---
---
+20
-10
+8.5
+11.5
+8.0
+12.0
+7.0
+13.0
+10
+3
+11.5
+14.5
+11.0
+15.0
+10.0
+16.0
+10
0
+8.5
+11.5
+8.0
+12.0
+7.0
+13.0
0
+3
+1.5
+4.5
+1.0
+5.0
0.0
+6.0
0
0
-1.5
+1.5
-2.0
+2.0
-3.0
+3.0
Table 7. Output Meter Accuracy (Option 002 with 003)
Measuring receiver indications
Test
(dBm)
instrument
OUTPUT
LEVEL
Meter
switch settings
indications
25 And 512 MHz
1000 MHz
(dBm)
(dB)
Min
Max
Min
Max
+20
-2
+15.0
+20.0
---
---
+20
-7
+10.0
+15.0
---
---
+20
-10
+6.5
+12.5
---
---
+10
+2
+9.0
+14.0
+8.5
+15.5
+10
0
+7.0
+12.0
+6.5
+13.5
0
+3
0.0
+5.0
-0.5
+6.5
0
0
-3.0
+2.0
-3.5
+3.5
13