TB 9-4931-495-24
link analyzers. The H80 option provides rear panel inputs that allow the TI to be swept
and frequency modulated by the outputs of the microwave link analyzer.
b. Time and Technique. The time required for this calibration is approximately 6
hours, using the microwave technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications, which
pertain to this calibration, are listed in table 1.
Stability
Power output
Model
10% line
10 dB power
Range
Accuracy
change
change
Leveled
Variation
8620A
Determined By Plug-In
and C
Cw, 7 MHz
20 kHz
20 kHz
0.35 dB
86210A
3 to 350
13 dBm
0.35 dB
Swp, 10 MHz
MHz
Cw, 10 MHz
20 kHz
20 kHz
0.5 dB, int
86220A-
10 to 1300
10 dBm
Swp, 15 MHz
H80
MHz
1 dB
Cw, 10 MHz
20 kHz
100 kHz
0.25 dB, int
86222A
0.01 to 2.4
13 dBm
Swp, 15 MHz
0.1 dB, ext
GHz
(20 mW)
1 dB
Cw, 10 MHz
20 kHz
100 kHz
0.25 dB, int
86222B
0.01 to 2.4
13 dBm
Swp, 15 MHz
0.1 dB, ext
GHz
(20 mW)
1 dB
Cw, 10 MHz
20 kHz
1 MHz
1.2 dB, int
86230B-
1.8 to 4.2
10 dBm
Swp, 15 MHz
0.1 dB, ext
H80
GHz
Cw, 30 MHz
30 kHz
1 MHz
0.7 dB, int
86241A-
3.2 to 6.5
4 dBm
Swp, 33 MHz
0.1 dB, ext
H80
GHz
Cw, 35 MHz
40 kHz
4 MHz
1 dB, int
86242A-
5.9 to 9
10 dBm
Swp, 40 MHz
H80
GHz
Cw, 40 MHz
40 kHz
2 MHz
1 dB, int
86250B-
8 to 12.4
8 dBm
Swp, 50 MHz
H80
Cw, 50 MHz
180 kHz
6 MHz
0.1 dB, ext
86250A-
12.4 to 18
7 dBm
Swp, 70 MHz
H80
GHz
(5 mW)
0.6 dB, int