Figure 1. Model 8620C - partial top view.
Figure 3. Model 8620C - partial top view.

Calibration Procedure For Sweep Oscillator Hewlett-Packard, Models 8620A And 86920C With Rf Plug-Ins, Models 86200 And 86300 Series; And Modules 8621A And 8621B Page Navigation
  3    4    5    6    7  8  9    10    11    12    13  
img
. . . .
TB 9-4931-495-24
Figure 2. Model 8620A - partial top view.
9. Full Sweep (Model 8620C)
a. Performance Check
NOTE
If necessary, zero TI using standard zeroing techniques.
(1) Select calibration scale with BAND selector.
(2) Connect multimeter between A2TP3 and A4 GND (fig. 3).
NOTE
On models prefixed with 1716, connect to chassis ground.