Equipment Setup
Table 4. Main Sweep Timing (Unmag)

Calibration Procedure For Delaying Time Base, Tektronix, Type 5B42; Time Base, Tektronix, Type 5B40; And Dual Trace Amplifier, Tektronix, Type 5A48 Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
img
. . . .
TB 9-4931-503-24
CAUTION
High  intensity  will  damage  oscilloscope  crt.
Adjust
INTENSITY control to lowest possible level consistent with
suitable display.
8. External Horizontal Gain
a. Performance Check
(1) Connect oscilloscope calibrator SOURCE/MEASURE CHAN 1 to TI EXT
HORIZ AMPL.
(2) Set oscilloscope calibrator VOLTAGE output for 0.5 V, 1 kHz.
(3) Adjust TI POSITION control to position dots to extreme right and left vertical
graticule lines.
(4) Adjust dual trace amplifier CH 1 POSITION control to vertically center the dots.
(5) Adjust oscilloscope calibrator EDIT FIELD knob for 10 divisions of horizontal
deflection on oscilloscope. If oscilloscope calibrator err display does not indicate within
3%, perform b below.
b. Adjustments Adjust oscilloscope calibrator EDIT FIELD knob for a 0.0% err
display indication. Adjust R52 EXT HORIZ GAIN (fig. 1) for a 10 division horizontal
deflection on oscilloscope (R).
Figure 1. Test instrument A board - left-side view.
6