• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Delayed Sweep and Delay Time/ Differential Time Measurement
Table 7. Delayed Sweep Timing (Mag) - Continued

Calibration Procedure For Delaying Time Base, Tektronix, Type 5B42; Time Base, Tektronix, Type 5B40; And Dual Trace Amplifier, Tektronix, Type 5A48 Page Navigation
  6    7    8    9    10  11  12    13    14    15    16  
img
. . . .
TB 9-4931-503-24
Table 6. Delayed Sweep Timing (Unmag)
Test instrument
Oscilloscope calibrator
err display
MAIN SEC/DIV
DLY'D SEC/DIV
MARKER
indications
switch settings
switch settings
output settings
( %)
s
s
S/D
3
10
10
10
s
s
S/D
3
20
20
20
s
s
S/D
3
50
50
50
0.1
ms
0.1
ms
0.1
mS/D
3
0.2
ms
0.2
ms
0.2
mS/D
3
0.5
ms
0.5
ms
0.5
mS/D
3
1
ms
1
ms
1
mS/D
3
2
ms
2
ms
2
mS/D
3
5
ms
5
ms
5
mS/D
3
10
ms
10
ms
10
mS/D
3
20
ms
20
ms
20
mS/D
3
50
ms
50
ms
50
mS/D
3
0.1
s
0.1
s
0.1
S/D
3
0.2
s
0.2
s
0.2
S/D
4
0.5
s
0.5
s
0.5
S/D
4
(7) Press SWP MAG pushbutton.
(8) Set TI switches and oscilloscope calibrator output as listed in first row of table 7.
NOTE
If necessary, adjust TI MAIN TRIG LEVEL control for a
stable display.
(9) Adjust TI POSITION control to align 2d time marker with 2d vertical graticule
line on display.
(10) Adjust oscilloscope calibrator EDIT FIELD knob to align 10th marker with 10th
vertical graticule line.  Oscilloscope calibrator err display will indicate within limits
specified in first row of table 7.
(11) Repeat technique of (8) through (10) above for remaining rows in table 7.
Table 7. Delayed Sweep Timing (Mag)
Test instrument
Oscilloscope calibrator
err display
MAIN SEC/DIV
DLY'D SEC/DIV
MARKER
indications
switch settings
switch settings
output settings
( %)
s
10
ns
10
nS/D
5
0.1
s
20
ns
20
nS/D
5
0.1
s
50
ns
50
nS/D
4
0.1
s
s
S/D
4
0.2
0.1
0.1
s
s
S/D
4
0.2
0.2
0.2
s
s
S/D
4
0.5
0.5
0.5
11


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.