External Horizontal Gain
Table 9. Sweep Timing (Mag)

Calibration Procedure For Delaying Time Base, Tektronix, Type 5B42; Time Base, Tektronix, Type 5B40; And Dual Trace Amplifier, Tektronix, Type 5A48 Page Navigation
  10    11    12    13    14  15  16    17    18    19    20  
img
. . . . . . .
TB 9-4931-503-24
Figure 3. Test instrument - adjustment locations.
15. Horizontal Time
a. Performance Check
(1) Connect oscilloscope calibrator SOURCE/MEASURE CHAN 1 to dual trace
amplifier CH 1 input.
(2) Press TRIGGERING SOURCE LEFT pushbutton.
(3) Set TI SECONDS/DIV switch and oscilloscope calibrator MARKER output as
listed in first row of table 8.
NOTE
If necessary, adjust TI LEVEL control for a stable display.
(4) Adjust TI POSITION control to align 2d marker with 2d vertical graticule line
on oscilloscope.
(5) Adjust oscilloscope calibrator EDIT FIELD knob to align 10th marker with 10th
vertical graticule line on oscilloscope. If oscilloscope calibrator err display indication is not
within limits specified in first row of table 8, perform b below.
(6) Repeat technique of (3) through (5) above for remaining rows in table 8.
Table 8. Sweep Timing (Unmag)
Oscilloscope calibrator
Test instrument
err display
SECONDS/DIV
MARKER output
indications
switch settings
settings
( %)
s
S/D
4
0.1
0.1
s
S/D
4
0.2
0.2
s
S/D
3
0.5
0.5
s
S/D
3
1
1
s
S/D
3
2
2
s
S/D
3
5
5
s
S/D
3
10
10
s
S/D
3
20
20
s
S/D
3
50
50
15