TB 9-4931-523-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification. This procedure provides instructions for the
calibration of Attenuators, Fixed and Variable (10 MHz to 26.5 GHz) (General). The
manufacturers' manuals were used as the prime data sources in compiling these
instructions. The equipment being calibrated will be referred to as the TI (test instrument)
throughout this bulletin.
a. Model Variations. Variations among models are described in appendix A.
b. Time and Technique. The time required for this calibration is approximately 2
hours using the dc and low frequency technique.
2. Forms, Records, and Reports. Forms, records, and reports required for calibration
personnel at all levels are prescribed by TB 750-25.
3. Calibration Description. TI parameters and performance specifications that pertain
to this calibration are listed in appendix A. TIs other than those listed in the appendix may
be checked using the techniques in this TB if the manufacturers' specifications are
available.
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 1 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set, AN/GSM-286; AN/GSM-287; or AN/GSM-440 and Secondary Reference
Calibration Standards Set, NSN 4931-00-621-7878. Alternate items may be used by the
calibrating activity. The items selected must be verified to perform satisfactorily prior to
use and must bear evidence of current calibration. The equipment must meet or exceed the
minimum use specifications listed in table 1. The accuracies listed in table 1 provide a four-
to-one ratio between the standard and TI. Appendix A is footnoted where the four-to-one
ratio may not be met.
5. Accessories Required. The accessories required for this calibration are common
usage accessories issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure.
Table 1. Minimum Specifications of Equipment Required.
Manufacturer and model
Common name
Minimum use specifications
(part number)
MEASURING RECEIVER
Frequency measurement:
Measuring receiver system N5531S
Range: 10 MHz to 26.5 GHz
consisting of: Spectrum Analyzer,
Agilent Model E4440A (E4440A), Power
Accuracy: 0.5%
meter, Agilent Model E4419B (E4419B),
or Power meter, Agilent Model N1911A
(N1911A) and Sensor modules, Agilent
Range: +10 dB to -113 dB
Model N5532A opt. 504 (504), 518 (518),
Accuracy: 0.125 dB
526 (526)