TB 9-4931-523-24
NOTE
Verify the proper Cal Factors are loaded for the power sensor
module being utilized.
NOTE
Ensure measuring receiver measurement accuracy is set to high.
7 a above and adjust output for 0.0 dBm.
(2) Configure measuring receiver for tuned RF level measurements and manually
enter carrier frequency into measuring receiver. Record reading as reference.
(3) Adjust signal generator output in 10 dBm steps to -70 dBm.
Allow the
measuring receiver to acquire the signal at each level.
NOTE
Use adapters as needed.
NOTE
Use 50 Ω to 75 Ω or 50 Ω to 93 Ω adapters when TI is 75 Ω or 93 Ω.
(4) Adjust signal generator output for 0.0 dBm indication on measuring receiver.
(5) Configure measuring receiver in the ratio mode.
(6) Connect the attenuator at the TI INSERTION POINT as shown in figure 1.
(7) Measure and record measuring receiver indication. Measured attenuation will
be within range and accuracies specified in appendix A for TI being calibrated.
NOTE
When calibrating a set of attenuators test each attenuator in
the set at each test frequency before setting a new
reference frequency.
(8) Repeat steps (1) through (7) above for remaining test frequencies recorded in
7 a above.
b. Adjustments. No adjustments can be made; however, a correction chart may be
prepared listing actual measuring receiver indications at frequencies of interest.
9. Variable Attenuation Measurement (10 MHz to 26.5 GHz)
a. Performance Check
NOTE
The frequency range of the TI will determine which sensor
module is used. 504: 100 kHz 4.2 GHz, 518: 10 MHz 18 GHz,
526: 30 MHz 26.5 GHz.
NOTE
Verify the proper Cal Factors are loaded into the controller for
the power sensor module being utilized.