TB 9-4931-523-35
APPENDIX C
(10) Press the RATIO key to deactivate, then press the GOLD (S) TUNED RF
LEVEL, 39.9 SPCL, 3.7 SPCL, and 1.9 SPCL keys.
(11) Note the power meter reading and record it as 'Power Meter Reading' in table C1
(12) Note the measuring receiver reading and record it as 'Init dB' reading in table
C1 alongside the
appropriate test frequency.
(13) Press the TI BLUE (SHIFT) key and then the SET REF (ZERO) key.
(14) Calculate the first cal point using the formula -40 + signal generator No. 1
output level Init dB reading. (Example -40 dBm + (6 dBm) - (-3 dBm) = -31 dBm). Record
this value, as 'First Cal Point' in table C1 alongside the appropriate test frequency.
(15) Set signal generator No. 1 to the level calculated in (14) above and press the
CALIBRATE key.
(16) Calculate the second cal point using the formula -80 + signal generator No.1
(No.2) output level 40 dBm to result of step (14). (Example -80 dBm + (6 dBm) - (-3 dBm)
= -71 dBm). Record this value, as 'Second Cal Point' in table C1 alongside the appropriate
test frequency.
(17) Set signal generator No. 1 to the level calculated in (16) above and press the
CALIBRATE key.
(18) Set signal generator No. 1 to the initial level of +6 dBm.
(19) Press 38.2 and SPCL keys and record the displayed value as 'First RF CAL
(20) Press 38.3 and SPCL keys and record the displayed value as 'Second RF CAL
(21) Press 38.4 and SPCL keys and record the displayed value as 'Set Ref CAL FAC'
(22) Remove TI from setup and repeat steps (4) through (21) for each of the test
(23) Press the FREQ and BLUE (SHIFT) AUTO TUNING keys.
(24) Set signal generator No. 1 to appropriate test frequency listed in table C1.
NOTE
For frequencies below 1.3 GHz proceed to step (27) below.
(25) Press 27.3 SPCL, enter the appropriate offset frequency from table C1, and
press the MHz key. (Ignore the displayed frequency reading.)
(26) Set signal generator No. 2 for an output of +8 dBm and the appropriate offset
(27) Press the GOLD(S) and TUNED RF LEVEL, 26.1 and SPCL keys.
C-3