TB 9-4931-523-35
NOTE
Software package USATA PDMAT () can be used in place of
steps 1a(1) through 1a(35) below, verifying that the TI meets
or exceeds accuracies listed in appendix A. Copies of the
software package are available from Commander, U. S. Army
Aviation and Missile Command, ATTN: AMSAM-TMD-LW,
Redstone Arsenal, AL 35898-5000.
1. Alternate Variable Attenuation Measurement (10 MHz to 18 GHz).
a. Performance Check
NOTE
When performing the following steps, if TI frequency range
extends above and below 1.3 GHz, it will be necessary to
perform two separate tests - one below and one above 1.3 GHz.
(1) Zero and calibrate power meter and measuring receiver RF power, then connect
(2) Record the 10 test frequencies selected in paragraph 7a in to the test frequencies
table C1 and
adjust signal generator No. 1 RF output
controls for +6
dBm.
NOTE
Use adapters as needed.
NOTE
Use waveguide to coaxial adapters when TI is waveguide.
NOTE
Use 50Ω-to-75Ω or 50Ω-to-93Ω adapters when TI is 75Ω or
93Ω.
(4) If the AUTO TUNING indicator light is off, press the measuring receiver FREQ
and BLUE (SHIFT) AUTO TUNING keys on the measuring receiver.
NOTE
For frequencies below 1.3 GHz proceed to step (8) below.