TB 9-4931-523-35
APPENDIX C
(28) Adjust signal generator No. 1 for a power meter indication as recorded in table
C1 as
'Power Meter
Reading.'
(29) Press 39.2 SPCL and enter the First RF Cal Fac from table C1 for the
appropriate frequency, and press the BLUE (SHIFT) % CAL FACTOR (MHz) keys.
(30) Press 39.3 SPCL and enter the Second RF Cal Fac from table C1 for the
appropriate frequency, and press the BLUE (SHIFT) % CAL FACTOR (MHz) keys.
(31) Press 39.4 SPCL and enter the Set Ref Cal Fac from table C1 for the
appropriate frequency, and press the BLUE (SHIFT) % CAL FACTOR (MHz) keys.
(32) Increase TI attenuation setting in one step increments (or one cardinal point as
desired.)
(33) Measure and record receiver system indication. Measured attenuation will be
within range and accuracies specified in appendix A for TI being calibrated.
(34) Repeat (32) and (33) above for remaining TI step increments.
(35) Repeat (23) through (34) above for remaining test frequencies recorded in table
b. Adjustments. No adjustments can be made; however, a correction chart may be
prepared listing actual receiver system indications at frequencies of interest.
C-4