TB 9-4931-530-50
SECTION I
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Differential Electronic Level Federal Products Corp., Model 232P-68. The
manufacturer's manual was used as the prime data source in compiling these instructions.
The equipment being calibrated will be referred to as the TI (test instrument) throughout
this bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 1
hour, using the physical technique.
2. Forms, Records, and Reports. Forms, records, and
reports required for calibration
personnel at all levels are prescribed by TB 750-25.
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Table 1. Calibration Description
Test instrument parameters
Performance specifications
Scale magnification
Range: 0 to 1000 arc s
Accuracy: 4/5 of dial division (40% FS)
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2
identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Reference Calibration
Standards Set NSN 4931-00-621-7878. Alternate items may be used by the calibrating
activity. The items selected must be verified to perform satisfactorily prior to use and must
bear evidence of current calibration. The equipment must meet or exceed the minimum use
specifications listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio
between the standard and TI.
5. Accessories Required. The
accessories listed in table 3
are issued as
indicated in
paragraph 4 above and
are used in this
calibration procedure. When necessary, these items
may be substituted by equivalent items, unless specifically prohibited.
2