TB 9-4931-535-35
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
Noise
Average reading
setting
(dBm)
30 kHz
<-100
10 kHz
<-104
3 kHz
<-108
1 kHz
<-111
300 Hz
<-115
100 Hz
<-122
30 Hz
<-127
10 Hz
<-132
3 Hz
<-137
Frequencies: Pwr line freq., 5 kHz, 100 kHz, 1 MHz, and 10 MHz
Accuracy: < -120 dBm
Spurious responses: < -80 dB
Spurious responses: < -80 dB
Accuracy: 20% of BW settings at 3 dB points
Selectivity (shape factor): <11:1 (60 dB BW/3 dB BW)
Flatness: 0.7 dB
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2 identifies the
specific equipment to be
used in
this
calibration procedure. This equipment is issued with Secondary Transfer Calibration Set
AN/GSM-287. Alternate items may be used by the calibrating activity when the equipment
listed in table 2 is not available. The items selected must be verified to perform
satisfactorily prior to use and must bear evidence of current calibration. The equipment
must meet or exceed the minimum use specifications listed in table 2.
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure.
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
LOW PASS FILTER
Frequency: 13.31 MHz
TLC 14-3EF (TLC 14-3EF)
Tektronix, Type SG505 (MIS-
Amplitude range: -20 dBm
30526/11)
Distortion: <90 dB
SYNTHESIZER/
Hewlett-Packard, Model
Amplitude range: -80 to +10 dBm
3335AOPT 001-K06 (MIS-
Accuracy: 0.075 dB
35958)
SECTION III