TB 9-4931-537-35
NOTE
The values listed in table 1, Limits (V) column, are calculated
for maximum resolution of appropriate range, for example a
100 mV input should be read on multimeter 100 mV range.
The first measurement for each perimeter may not auto
range properly. If this occurs set multimeter to appropriate
range, then set back to auto range.
(5) Set calibrator to STANDBY, then press RESET pushbutton.
e. Ac Voltage. Calibrator, John Fluke, Model 5700A/( ) Multimeter, Hewlett
Packard, Model 3458A; and Amplifier, John Fluke, Model 5725A/( ).
CAUTION
(1) Press multimeter FUNCTION/RANGE ACV pushbutton; then set calibrator
to (m)V, (k)Hz outputs listed in table 1. Record multimeter indications in space
provided.
(2) Set calibrator to STANDBY, then press RESET pushbutton.
f. Dc Current. Calibrator, John Fluke, Model 5700-A/( ), Multimeter, Hewlett
Packard, Model 3458A and Amplifier, John Fluke, Model 5725A/( ).
7