TB 9-4931-537-35
CALIBRATOR
OUTPUT SENSE
HI
HI
LO
LO
SENSE INPUT
HI
HI
GUARD GROUND
LO
LO
CURRENT
GUARD
I
OUTPUT
HI
LO
AMPLIFIER
(3) Set calibrator for a 0 Ω output. Set calibrator's EX SENS to on (illuminated).
Set 2 Wire Comp softkey to OFF.
(4) Multimeter indication will be 00.00000 ( 8 digits). Record indication in table
(5) Set calibrator to (k, M) Ω outputs listed in table 1. At each resistance output,
adjust the calibrator output adjustment control knob for a calibrator control display
reading equal to multimeter indication. The calibrator control ERROR display
indication will be within specified limits of table 1. Record calibrator ERROR display
indication in space provided.
(6) Set calibrator to STANDBY, then press RESET pushbutton.
9