TB 9-4931-540-35
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
1 MΩ flatness
Flatness: 0.7 dB
Flatness: 1.5 dB
Average
Noise
Resolution
indication
(dBm)
setting
30 kHz
<-100
10 kHz
<-104
3 kHz
<-108
1 kHz
<-111
300 Hz
<-115
100 Hz
<-122
30 Hz
<-127
10 Hz
<-132
3 Hz
<-137
40 Hz <-123 dBm (1 Hz)
Frequencies: Pwr line freq., 5 kHz, 100 kHz, 1 MHz, and 10
MHz
Accuracy: <-120 dBm
Spurious responses: <-80 dB
Spurious responses: <-80 dB for 50Ω input
<-70 dB for 1 MΩ input
Frequency range: 3 Hz to 30 kHz
Accuracy: 20% of BW settings at 3 dB points
Selectivity (shape factor): <11:1 (60 dB BW/3 dB BW)
Flatness: 0.7 dB
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2
identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration Set
AN/GSM-287. Alternate items may be used by the calibrating activity. The items selected
must be verified to perform satisfactorily prior to use and must bear evidence of current
calibration. The equipment must meet or exceed the minimum use specifications listed in
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure.