TB 9-4931-540-35
(g)
ENTRY REFERENCE LEVEL to -50 dBm.
(h)
RBW-VBW-ST ENTRY RES BW to 3 Hz.
(i)
RBW-VBW-ST ENTRY VIDEO BW to 1 Hz.
(j)
ENTRY SAVE and then 1.
(k)
ENTRY CF STEP SIZE to 60 Hz.
(l)
SWEEP ENTRY MANUAL.
(m) ENTRY STEP ⇓.
(5) Average TI marker amplitude indication will be <-80 dB.
(6) Press ENTRY STEP ⇓ key. Average TI marker amplitude indication will be
<-80 dB.
(7) Repeat (6) above three times.
(8) Press ENTRY STEP ⇑ key six times. Average TI marker amplitude indication
will be <-80 dB.
(9) Press ENTRY STEP ⇑ key. Average TI marker amplitude indication will be
<-80 dB.
(10) Repeat (9) above three times.
(11) Press keys and enter values using ENTRY keys as listed in (a) through (d)
below:
(a) ENTRY RECALL and then 1.
(b) ENTRY CF STEP SIZE to 5 kHz.
(c) SWEEP ENTRY MANUAL.
(d) ENTRY STEP ⇓.
(12) Repeat (5) through (10) above.
(13) Press keys and enter values using ENTRY keys as listed in (a) through (d)
below:
(a) ENTRY RECALL and then 1.
(b) ENTRY CF STEP SIZE to 100 kHz.
(c) SWEEP ENTRY MANUAL.
(d) ENTRY STEP ⇓.
(14) Repeat (5) through (10) above.
(15) Press keys and enter values using ENTRY keys as listed in (a) through (d)
below:
(a) ENTRY RECALL and then 1.
(b) ENTRY CF STEP SIZE to 1 MHz.
(c) SWEEP ENTRY MANUAL.
(d) ENTRY STEP ⇓.
19