TB 9-4931-540-35
(b) RBW-VBW-ST ENTRY RES BW to 30 kHz.
(c) SWEEP ENTRY MANUAL to 300 kHz.
(11) Average TI marker amplitude indication will be <-100 dBm.
(12) Repeat technique of (10)(b), (c), and (11) above for remaining TI settings and
Average marker
SWEEP ENTRY
RBW-VBW-ST
amplitude
MANUAL
ENTRY RES BW
indications
settings
settings
(dBm)
(kHz)
10
kHz
100
<-104
3
kHz
30
<-108
1
kHz
10
<-111
300
Hz
10
<-115
100
Hz
10
<-122
30
Hz
10
<-127
10
Hz
10
<-132
3
Hz
10
<-137
(13) Press keys and enter values using ENTRY keys as listed in (a) through (e)
below:
(a)
ENTRY blue shift and then ENTRY CAL OFF.
(b)
SWEEP ENTRY MANUAL to 40 Hz.
(c)
RBW-VBW-ST ENTRY RES BW to 3 Hz.
(d)
ENTRY blue shift and then ENTRY CAL ON.
(e)
MARKER/CONTINUOUS ENTRY NOISE LVL to on.
NOTE
Wait until (1 Hz) is displayed behind TI marker amplitude
dBm indication before proceeding to (14) below.
(14) Average TI marker amplitude indication will be <-123 dBm.
15. Low Frequency
Response
a. Performance Check
(1) Ensure TI front panel inputs are open.
NOTE
60 Hz represents power line frequency; if power line frequency
is different than 60 Hz, use current power line frequency in
(2)(g) and (h) below.