TB 9-4931-541-50
GIMBALED FIXTURE
MOUNTING THUMBSCREW
CLAMPING BLOCK
LOCKING SCREW
TO TI INDICATOR
TO LENGTH GAUGE
X1 CONNECTOR
INDICATOR
X1 CONNECTOR
c. Connect both indicators to 115 V ac power source. Turn power switches to on and
allow length gauge and TI to warm up for 15 minutes.
d. Both indicators will indicate ENT ...CL, press CL. The last three digits on the
readout may continually fluctuate depending on the environmental conditions and the
warm-up time allowed. The ACTL and INCH LEDs should be activated.
NOTE
If ACTL is not activated on the display, continue pressing
MOD until ACTL flashes on display, then press ENT and CL.
If any other indications are incorrect, see PG 7 paragraph 4.6.4
of the Dial Indicator Calibrator Itall Model 700 manual for
proper programming.
4