TB 9-4931-700-50
b. Connect autotransformer to
115-volt ac
(alternating current) and
adjust its output
control to obtain a 115-volt output.
c. Set LINE switches to ON and allow 1 hour for equipment to warm-up and
stabilize.
SECTION III
NOTE
not within tolerance, perform the power supply check in
After
making adjustments. Do not perform power supply checks if
all other parameters are within tolerance.
a. Performance Check
(1) Connect TI front panel output to input of frequency counter (A5), using cable
(B8).
(2) Position TI controls as listed in (a) through (d) below:
(a) Press FREQUENCY RANGE X10 pushbutton.
(b) Adjust FREQUENCY control to obtain FREQ scale indication of 1.
(c) Press VOLTAGE RANGE 1V pushbutton.
(d) Set voltage controls to obtain indication of 1.000000 V.
(3) If frequency counter does not indicate between 98.0392 and 102.041 ins
(milliseconds), perform b below.
(4) Repeat technique of (2) and (3) above, using settings and indications listed in