TB 9-4933-220-50
ANALYZER TEST SET XM-28
8. DC Volts and Tracking
standard does not indicate between 0.95 and 1.05 volts,
perform 8b(1) and (2) below.
a. Performance Check.
(3) Release TI METER FINE switch.
(1) Connect DC voltage standard (A1) to TI,
connecting plus terminal to E22 (fig 1) and minus
(4) Disconnect lead from E22 and connect to
terminal to E19 (fig 1), using two adapters and two leads
(B1 through B4).
(5) Repeat technique of (2) above at TI switch
(2) Set and hold METER switch to FINE
positions and meter indications listed in table 5. DC
position and adjust DC voltage standard output voltage
voltage standard will indicate within limits specified.
until TI VOLTS meter indicates 1 volt. If DC voltage
Figure 1. Analyzer test set - - rear view.
7