TB 9-4935-365-24
f. Remove short. For model 8120A, press RANGE 1 pushbutton.
a. Performance Check
(1) Connect TI INPUT HI to calibrator OUTPUT HI and TI INPUT LO to
calibrator OUTPUT LO.
(2) Set calibrator for an output amplitude of 1.0000 Vdc. If TI does not indicate
within limits specified in first row of table 3, perform corresponding adjustments in table 3.
(3) Repeat technique of (2) above, using calibrator outputs and TI range settings
listed in table 3. If TI does not indicate within limits specified, perform corresponding