TB 9-4935-365-24
a. Performance Check
(1) Press FUNCTION MA-DC and RANGE 1 pushbuttons.
(2) Set calibrator for an output amplitude of 1 mA dc. If TI does not indicate within
limits specified in first row of table 5, perform corresponding adjustment in table 5.
(3) Repeat technique of (1) and (2) above using settings listed in table 5.
TI will
indicate within limits specified in table 5.
Calibrator
Test instrument
Adjustment
Range
Indication limits
Output
setting (MA)
Min
Max
(R)
1 mA
1
0.9988
1.0012
MA
100 A
1
100 A
99.88
100.12
1 0 mA
10
9.988
10.012
01 mA
0
100
99.88
100.12
A
1000
998.8
1001.2
(4) Set calibrator output to minimum.
b. Adjustments. No further adjustments can be made.
a. Performance Check
(1) Press function 10 M
pushbutton.
(2) Set calibrator for a 10 M
nominal output.
(3) Rotate calibrator knob below EDIT FIELD pushbutton to adjust calibrator
display indication to equal TI indication. If calibrator err display does not indicate within
limits specified in first row of table 6, perform corresponding adjustments in table 6.
(4) Repeat technique of (1) through (3) above, using outputs and settings listed in
table 6. If calibrator err display does not indicate within limits specified in table 6, perform
Test instrument
Calibrator
Adjustments
err display
Range pushbutton
Nominal output
indication (%)1
(R)
settings
0.06 (0.11)
10 M
10 M
10 MEG CAL (10 MEG)
0.03 (0.06)
100 (100k)
100 k
K OHM CAL (K OHMS)
0.03 (0.06)
1 (1 k)
1k
0.03 (0.06)
10 (10 k)
10 k
0.03 (0.06)
1
1000 (1 M)
1M
Values in parenthesis apply to model 8120A.
b. Adjustments. No further adjustments can be made.