SECTION III. CALIBRATION PROCESS

Calibration Procedure For Torque Wrench Weinschel, Model 2098-5065 Page Navigation
  1    2  3
img
. . . .
TB 9-5120-204-24
(1) Attach TI to torque tester, using adapters as required.
(2) Slowly apply torque to TI and torque tester in a cw direction until audible "click"
is heard. (Do not over-torque torque tester).
(3) Slowly release torque until only tare torque of TI is applied.
(4) Repeat (2) and (3) above two more times.
(5) Remove TI from torque tester.
8. Torque
a. Performance Check
(1) Note torque value setting on handle of TI. If no value is stamped on handle, use
value in table 1.
(2) Insert TI in torque tester, using adapters as required. Insert in horizontal position.
(3) Gradually apply perpendicular force on handle of TI until audible "click" is
heard. (Do not over-torque torque tester). Note indication on torque tester.
(4) Slowly release torque until only tare torque of TI is applied.
(5) Repeat (3) and (4) above two more times.
(6) Compute average reading of indications in (3) and (5) above. If average reading
is not within 4% of torque value noted in (1) above, perform b below.
b. Adjustments. Adjust in accordance with manufacturer's instruction manual.
9. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.