SECTION III. CALIBRATION PROCESS
Table 3. Calibration Points

Calibration Procedure For Torque Wrench Power-Dyne, Model Ph-602 Page Navigation
  1    2  3  4    5    6  
img
. .
TB 9-5120-206-24
TI
DRIVE BAR
SOCKET ADAPTER
TORQUE CELL
TORQUE MOUNTING ASSEMBLY
MOUNTING PLATE
FORCE/TORQUE INDICATOR
Figure 1. Torque - equipment setup.
e. Connect torque cell no. 1 to force/torque indicator, using cable supplied with
force/torque indicator.
f. Connect force/torque indicator to appropriate power source. Turn power switch to
ON and allow the unit to warm-up for 30 minutes. Select channel 3 using the CHANNEL
+ and keys.
NOTE
Whenever direction of torque is changed (cw or ccw),
force/torque indicator setup must be changed and TI must be
exercised three times.
soft keys.
g. Press the F4 key on the force/torque indicator until you see the
Press the
soft keys as necessary to select torque cell no. 1 and ccw direction. Insure
the serial number matches the torque cell being used.
h. Press the F4 key on the force/torque indicator until you see the UNIT soft key.
Press the UNIT soft key as necessary to select ft-lb.