Torque

Calibration Procedure For Torque Wrench Power-Dyne, Model Ph-602 Page Navigation
  1    2    3    4    5  6
img
. . . .
Figure 2. Calibration adjustments.
(l)  Press the F4 key on the force/torque indicator until you see the UNIT soft
key. Press the UNIT soft key as necessary to select ft-lb.
to select the GROSS mode of operation. Press the
(m) Press the SIGNAL
F4 key on the TI indicator until you see the Acal soft key. Press the Acal soft key as
necessary to activate the Acal enunciator.
(n) Connect equipment and repeat (1) through (6) above. If TI meter indication
is not within 3 percent of force/torque indicator indication, repeat (7) above (R).
(o) Repeat paragraphs 7 e thru 7 i and 8 a above.
9. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
7