TB 9-5210-208-35
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Length measurement
Range:
Vernier
Class 1: 0 to 48 in.
Class 2: 0 to 900 mm
Class 3: 0 to 36 in. and 0 to 900 mm
Accuracy: 0.001 in. per each 6 in. of beam length
0.02mm per each 300mm of beam length
Length measurement
Accuracy
0.001 in per each 6 in of beam length.
Dial
0.02mm per each 150mm of beam length
Length measurement
Accuracy
0.001 in per each 6 in of beam length
Digital
0.02mm per each 300mm of beam length
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286. Alternate items may be used by the calibrating activity
when the equipment listed in table 2 is not available. The items selected must be
verified to perform satisfactorily prior to use and must bear evidence of current
calibration. The equipment must meet or exceed the minimum use specifications listed
5. Accessories Required. The
accessories required for this
calibration are
common
usage accessories issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure.
Table 2. Minimum Specifications of
Equipment Required
Common name
Minimum use specifications
Manufacturer and model (part number)
GAGE BLOCKS
Range: 0 to 48 in
(7915946)
(8 PIECE SET)
Accuracy: Grade 3
GAGE BLOCK
Range: 0.050 to 4 in.
(7901961)
(81 PIECE SET)
Accuracy: Grade 3
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and are
preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.