TB 9-5210-214-35
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification. This bulletin
provides instructions for the
calibration of Dial Indicators Transfer Level. The manufacturer's instruction manual
was used as the prime data source in compiling these instructions. The dial indicators
will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations . Variations among models are described in text.
b. Time and Technique . The time required for this calibration is approximately
0.5 hour for each TI using the physical technique.
2. Forms,
Records, and
Reports
a.
Forms, records and
reports required for
calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Style1
A
Spindle, parallel to dial
Dovetail
Mounting ridge perpendicular to dial
Type1
1
1/10000 (0.0001) inch graduations
2
5/10000 (0.0005) inch graduations
3
1/1000 (0.001) inch graduations
4
5/100000 (0.00005) inch graduations
1 graduation at any point within the range of the indicator except type 1
Accuracy
long range indicators (more than 2 1/3 revolutions) shall be 1 graduation
within the first 2 1/3 revolutions and 5 graduations at any point beyond 2
1/3 revolutions
1This
specification is for information only and is not verified in this bulletin.