Figure A-2. Unserviceable example of DA Form 3023.

Calibration Procedure For M2 Headspace And Timing Gage Set Model 5351217 Page Navigation
  5    6    7    8    9  10
img
.
TB 9-5220-213-24
Recorded Readings Worksheet
5351211
5351211
5351213
5351214
HEADSPACE
HEADSPACE
TIMING
TIMING
FIRE
GO
NO GO
NO FIRE
0.2015 - 0.2020
0.2060 - 0.2065
0.1140 - 0.1160
0.0200 0.0220
Actual readings
taken
9 for Headspace
6 for timing
____________
____________
____________
____________
____________
____________
TOTAL
9
9
6
6
Divided by # of
readings taken
Average
To be entered on
3023 card in
ACTUAL GAGE
DIMENSIONS
Figure A-3. Worksheet.
NOTE
If any single reading is not within the gage's range, this single
out of tolerance value will be annotated on the DA 3023 card
instead of the average reading.  The gage is unserviceable.
A-5