Boresight Collimator Test Set Accuracy - Continued
Figure 11. Stationary reticle reflected image coarse adjustment.

Calibration Procedure For Test Set, Boresight Collimator, Ts-3784/Tas Page Navigation
  11    12    13    14    15  16  17    18    19    20  
img
. . . . . . . . . . . . . . .
TB 9-5855-1891-40
b. Adjustments
(1) While viewing through EYEPIECE (fig. 6) look for BRIGHT GREEN DOT (fig. 5).
(2) Loosen COARSE ADJUSTMENT SCREW LOCKNUT and adjust COARSE
ADJUSTMENT SCREW to center BRIGHT GREEN DOT on reflection of your eye (fig. 5).
(3) While viewing through EYEPIECE, turn COARSE ADJUSTMENT SCREW
(fig. 5) until STATIONARY RETICLE REFLECTED IMAGE (fig. 6) is in field-of-view.
(4) While viewing through EYEPIECE turn COARSE ADJUSTMENT SCREW
(fig. 5) to adjust elevation setting and azimuth setting of STATIONARY RETICLE
REFLECTED IMAGE (fig. 6) to read between 4 arc-minutes and 6 arc-minutes on
NUMBER MARKINGS (fig. 6) (R).
(5) Tighten COARSE ADJUSTMENT SCREW LOCKNUT (fig. 5).
(6) Loosen two screws on COARSE ELEVATION ECCENTRIC PIVOT (fig. 11).
NOTE
Make  all  adjustments  slowly  to  ensure  STATIONARY
RETICLE REFLECTED IMAGE passes through FIELD-OF
VIEW (fig. 11).
(7) Loosen COARSE ELEVATION ECCENTRIC PIVOT LOCKNUT (fig. 11).
(8) Look through EYEPIECE, alternately adjust AZIMUTH STOP SCREW (fig. 12),
and adjust COARSE ELEVATION ECCENTRIC PIVOT until STATIONARY RETICLE
REFLECTED IMAGE is seen in FIELD-OF-VIEW but not on top of STATIONARY
RETICLE (fig. 11) (R).
(9) Hold COARSE ELEVATION ECCENTRIC PIVOT while tightening COARSE
ELEVATION ECCENTRIC PIVOT LOCKNUT (fig. 11).
17