SECTION I. IDENTIFICATION AND DESCRIPTION
Figure 1. Thermal sight (TI) with top cover of case and shroud removed.

Calibration Procedure For Thermal Sight Collimator, Sm-D-969722 Page Navigation
  1  2  3    4    5    6    7  
img
. . . . . . . . .
TB 9-5855-1892-24
Table 2. Minimum Specifications of Equipment Required
Minimum use
Manufacturer and model
Common name
specifications
(part number)
Range: 0 to 5 C differential
DIFFERENTIAL
Omega Engineering 0.020 Type K
Accuracy: 0.11 C
(7917040)1
THERMOCOUPLE SET
Hewlett Packard, Model 3458A
MULTIMETER
Range: 0 to 15 V
2
(3458A)
Accuracy:
1Part
of calibration kit 7917034, limited deployed.
Combined accuracy of thermocouples and multimeter is 4.5 V.
2
Table 3. Accessories Required
Description
Common name
(part number)
COTTON TIP APPLICATOR1
Swab with cotton tip on one end
HOLDING CLAMP2
Mounting post clamp (7917039) 3
LABORATORY CLAMP2
Laboratory clamp (MS-36012-1) 3
PASTE REMOVAL SOLVENT1
WD40 or equivalent
THERMOCOUPLE MOUNTING POST2
Mounting post with base (7917037)3
THERMALLY CONDUCTIVE PASTE1
Omegatherm 201 or equivalent
1Additional
equipment required.
2Two
required.
3Component part of calibration kit 7917034, limited deployed.
SECTION III
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6, 7 and 8 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in tables 2 and 3.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure.  Additional
maintenance information is contained in TM 9-4931-586-30&P for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
7. Preliminary Checks
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance check where applicable.
3