Figure 1. Thermal sight (TI) with top cover of case and shroud removed.
Equipment Setup

Calibration Procedure For Thermal Sight Collimator, Sm-D-969722 Page Navigation
  1    2    3  4  5    6    7    8    9  
img
. . . . . .
TB 9-5855-1892-24
c. Remove four screws from large cover plate and remove large cover plate (fig. 2).
d. Set ON/ALIGN switch to ALIGN (fig. 3).
SMALL COVER
PLATE
ON/ALIGN
MRT
SWITCH
BAR
ON
X
ALIGN
Figure 3. Large cover plate removed (top view).
e. Loosely replace large cover plate (fig. 2).
f. Perform steps (1) through (6) below:
(1) Ensure TI ON/OFF switch is in OFF position.
(2) Position TI 1 C/4.5 C/10 C switch to 4.5 C.
(3) Connect ac power cable supplied to J1 (fig. 4).
NOTE
TI should be allowed to stabilize in calibration environment for
a minimum of 1 hour.
Figure 4. TI side view.