TB 9-5855-1892-24
TARGET SELECTOR
MRT
SWITCH
BAR
ON
X
ALIGN
TARGET SELECTOR
SWITCH SCREWS
Figure 7. Loosening target selector switch mount. (ON/ALIGN switch is shown for clarity only).
(11) Increase pressure on thermocouple No. 2 to produce a slight bow in
thermocouple wire by sliding laboratory clamp forward (toward TI) in holding clamp, then
(12) Repeat technique of steps 9 a (1) through 9 a (3) (e) and record values on "MRT
Target Temperature Worksheet". Temperature error will be between 0.45 and 0.45 C; if
not, perform b (2) below and repeat this step.
Figure 8. Repositioning TC2 thermocouple wire to MRT target housing.
(10) Loosen holding clamp (fig. 5) slightly.
10