Temperature Check
Temperature Check - Continued

Calibration Procedure For Thermal Sight Collimator, Sm-D-969722 Page Navigation
  4    5    6    7    8  9  10    11    12    13  
img
. . . .
TB 9-5855-1892-24
TARGET SELECTOR
MRT
SWITCH
BAR
ON
X
ALIGN
TARGET SELECTOR
SWITCH SCREWS
Figure 7. Loosening target selector switch mount. (ON/ALIGN switch is shown for clarity only).
(11) Increase pressure on thermocouple No. 2 to produce a slight bow in
thermocouple wire by sliding laboratory clamp forward (toward TI) in holding clamp, then
tighten holding clamp (fig. 5).
(12) Repeat technique of steps 9 a (1) through 9 a (3) (e) and record values on "MRT
Target Temperature Worksheet". Temperature error will be between 0.45 and 0.45 C; if
not, perform b (2) below and repeat this step.
Figure 8. Repositioning TC2 thermocouple wire to MRT target housing.
(10) Loosen holding clamp (fig. 5) slightly.
10