TB 9-5855-1892-24
NOTE
A temperature correction greater than limits specified in table
4 is possible. A temperature correction is added or removed
each time technique of 9 b (2) is performed depending on dip
switch settings. Repeating technique of 9 b (2) will cause
additional temperature corrections each time the step is
performed.
Figure 9. Example of Dip Switch correction setting for 0.025 to 0.075 C.
(e) Perform steps 1 through 3 below:
1 Set TI ON/OFF switch to ON, and wait until READY lamp illuminates
and remains illuminated.
2
Set ON/ALIGN switch to ALIGN, wait 10 seconds and set ON/ ALIGN
switch to ON.
3 Verify temperature error correction has been stored by observing READY
lamp blinking at 2 Hz rate.
Table 4. Temperature Correction Settings
Temperature error
Dip switch
Dip switch
C
1
2
A
B
C
D
+0.375 and greater
On
On
On
Off
+0.325 to +0.375
Off
On
On
Off
+0.275 to +0.325
On
Off
On
Off
+0.225 to +0.275
Off
Off
On
Off
+0.175 to +0.225
On
On
Off
Off
+0.125 to +0.175
Off
On
Off
Off
+0.075 to +0.125
On
Off
Off
Off
12