TB 9-5885-1892-35
5. Accessories Required. Common usage accessories not
listed in this
calibration
procedure and peculiar accessories listed in table 3 are issued as indicated in paragraph 4
above.
Table 2. Minimum Specifications of
Equipment Required
Minimum use
Manufacturer and model
Common name
specifications
(part number)
DIFFERENTIAL
Omega Engineering 0.020 Type K
Range: 0 to 5 C differential
THERMOCOUPLE SET
(7917040)1
Accuracy: 0.11 C
Range: 0 to 15 V
Hewlett Packard, Model 3458A
Accuracy:
(3458A)
2
1Part
Combined accuracy of thermocouples and multimeter is 4.5 V.
2
Description
Common name
(part number)
COTTON TIP APPLICATOR1
Swab with cotton tip on one end
HOLDING CLAMP2
Mounting post clamp (7917039) 3
LABORATORY CLAMP2
Laboratory clamp (MS-36012-1) 3
PASTE REMOVAL SOLVENT1
WD40 or equivalent
THERMOCOUPLE MOUNTING POST2
Mounting post with base (7917037)3
THERMALLY CONDUCTIVE PASTE1
Omegatherm 201 or equivalent
1Additional
equipment required.
2Two
required.
3Component part of calibration kit 7917034, limited deployed.
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6, 7 and
8 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in TM 9-4931-586-30 & P for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.