TB 9-5855-1892-35
Temperature error
Dip switch
Dip switch
1
2
C
A
B
C
D
+0.375 and greater
On
On
On
Off
+0.325 to +0.375
Off
On
On
Off
+0.275 to +0.325
On
Off
On
Off
+0.225 to +0.275
Off
Off
On
Off
+0.175 to +0.225
On
On
Off
Off
+0.125 to +0.175
Off
On
Off
Off
+0.075 to +0.125
On
Off
Off
Off
+0.025 to +0.075
Off
Off
Off
Off
Off
On
On
On
-0.025 to -0.075
On
Off
On
On
-0.075 to -0.125
Off
Off
On
On
-0.125 to -0.175
On
On
Off
On
-0.175 to -0.225
Off
On
Off
On
-0.225 to -0.275
On
Off
Off
On
-0.275 to -0.375
Off
Off
Off
On
-0.325 and less
(e) Perform steps 1 through 3 below:
1 Set TI ON - OFF switch to ON, and wait until READY lamp illuminates
and remains illuminated.
2 Set ON ALIGN switch to ALIGN, wait 10 seconds and set ON ALIGN
switch to ON.
3 Verify temperature error correction has been stored by observing READY
lamp blinking at 2 Hz rate.
(f) Perform steps 1 and 2 below:
1
Set TI ON - OFF switch to OFF.
CAUTION
Disconnect ac power cable from TI before changing dip switch
settings and reconnect TI ac power cable after changing dip
switch settings.
2 Set DIP SWITCH 1, A and B to on and DIP SWITCH 2, C and D to on,
then loosely replace LARGE COVER PLATE (fig. 2).
a. Deenergize and disconnect all equipment.
b.
Remove TI LARGE
COVER PLATE
(fig. 2), place TARGET
SELECTOR SWITCH
back in original position and tighten TARGET SELECTOR SWITCH SCREWS (fig. 7).