TB 9-5855-1892-35
MRT TARGET
HOUSING
THERMOCOUPLE WIRE ON FLAT SURFACE
IN THERMALLY CONDUCTIVE PASTE
TC2
HEATER PLATE
(INSIDE)
TC1
LABORTORY
CLAMP
Figure 8. Repositioning TC2 thermocouple wire to MRT
target housing.
(10) Loosen HOLDING CLAMP (fig. 5) slightly.
(11) Increase pressure on thermocouple No. 2 to produce a slight bow in
thermocouple wire by sliding LABORATORY CLAMP forward (toward TI) in HOLDING
(12) Repeat technique of steps 9a(1) through 9a(3)(e) and record values on "MRT
Target Temperature Worksheet". Temperature error will be between -0.45 and +0.45 C; if
not, perform b(2) below and repeat this step.
(13) Deenergize and disconnect all leads.
b. Adjustments
(1) Perform steps (a) through (f) below:
(a)
Set digital multimeter power to ON.
(b)
Set digital multimeter to measure dc volts and auto range.
(c)
Verify digital multimeter indications vary between 0 2 V dc and 15 2 V dc.
(d)
(e) Wait for READY lamp to stop blinking and remain illuminated.
11