TB 9-5975-219-24
T
TI hyper-
Hyper-terminal
Resistance
d Test
terminal
est instrument indications
S
standard
( )
escription
range
ettings
(pin numbers)
setting
(k )
Min
Max
command
S18
1.5
1425
1575
TK ID Resistor (J2E_V)
S19
1.5
1425
1575
TK ID Resistor (J3E_V)
S20
1.5
1425
1575
TK Resistance (J4A_B)
S26
5
4750
5250
S31
50
47500
52500
S37
100
95000
105000
(5) Set calibrator to standby, and disconnect equipment setup.
b. Adjustments. If TI fails any parameter, perform Field-Calibration in Section IV.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
ALIGNMENT PROCESS
NOTE
Field-calibration should be performed only if TI fails any
parameter in the performance verification. The entire Field-
Calibration should be documented as a reportable adjustment.
a. Ensure that the TI is connected to the calibrator and instrument controller as
described in paragraph 7 a through 7 i above.
b. Connect multimeter HI and LO terminals to the ATE SINGLE POINT + and -
banana connectors located on the side of the ATE calibrator.
c. Select the HyperTerminal window, and at the ? prompt, enter Y on the instrument
controller keyboard. Do not press Enter until the instruction on the HyperTerminal
window has been completed.
d. Press Enter and follow the instruction displayed on the instrument controller
HyperTerminal window.