• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
APPENDIX C. THREE PORT CIRCULATOR TEST INSTRUMENT IDENTIFICATION
APPENDIX E. ALTERNATE CALIBRATION PROCESS FOR DIRECTIONAL COUPLERS (10MHZ TO 18GHZ)

Calibration Procedure For Directional Couplers, Vswr Bridges, Coaxial Circulators, Power Splitters, And Power Dividers (10 Mhz To 40 Ghz) (General) Page Navigation
  30    31    32    33    34  35  36    37    38    39    40  
img
. .
TB 9-5985-314-24
APPENDIX D
POWER SPLITTERS AND POWER DIVIDERS
TEST INSTRUMENT IDENTIFICATION
OUTPUT PORT
Model/part
Frequency range
Insertion loss
(GHz)1
TRACKING (dB)2
number
Manufacturer
(dB nom)
K241 ( )3
Wiltron
Dc
to
40
Dc
to
6
7.0
0.3
6
to
18
7.5
0.3
18
to
26.5
8.0
0.6
26.5  to
40
8.5
0.6
64
PS018
Weinschel
Dc
to
4
<0.15
4
to
8
<0.20
8
to
18
<0.25
1506A
Weinschel
Dc
to
18
6 dB, -0.2, +1.2 dB
max to 10.0 GHz;
Dc
to
4
0.2
+1.5 dB max to
4
to
10
0.4
18.0 GHz
10
to
18
0.5
1870A3
Weinschel
Dc
to
18
6 dB +1.5
Dc
to
8
0.15
8
to
18
0.2
64
11667A
Hewlett-Packard
Dc
to
4
<0.15
4
to
8
<0.20
8
to
18
<0.25
11667B3
64
Hewlett-Packard
Dc
to
26.5
Dc
to
18
<0.25
18
to
26.5
<0.40
1Not
checked below 10 MHz.
2Between
output ports.
3Attach test report for the frequencies requested or required by customer.
4Value determined in calibration process.
D-1/(D-2 Blank)