Alternate Calibration Process for Directional Couplers (10 MHz to 18 GHz) - Continued
APPENDIX F. ALTERNATE CALIBRATION PROCESS FOR VSWR BRIDGES (10MHZ TO 18GHZ)

Calibration Procedure For Directional Couplers, Vswr Bridges, Coaxial Circulators, Power Splitters, And Power Dividers (10 Mhz To 40 Ghz) (General) Page Navigation
  35    36    37    38    39  40  41    42    43    44    45  
TB 9-5985-314-24
NOTE
When determining directivity, in- or out-of-tolerance condition,
all of the below must be considered:
1. Termination  mismatch  errors  can  cause
measured
directivity to appear slightly lower than normal.
2. A 10 percent variation in directivity is acceptable for most
directional couplers used in direct support of field activities in
the U. S. Army.
3. A directional coupler may be out of tolerance at a specific
frequency and still be useable over the rest of its range.
2. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
E-5/(E-6 Blank)