Alternate Calibration Process for VSWR Bridges (10 MHz to 18 GHz) - Continued
Final Procedure

Calibration Procedure For Directional Couplers, Vswr Bridges, Coaxial Circulators, Power Splitters, And Power Dividers (10 Mhz To 40 Ghz) (General) Page Navigation
  39    40    41    42    43  44  45    46    47    48    49  
img
. . .
TB 9-5985-314-24
Table F2. Return Loss Measurements (with N-Short)
Test
Mismatch
Mismatch
Mismatch
Mismatch
frequency
2334-001-1
2334-001-2
2334-001-3
2334-001-4
(33) Repeat steps (21) through (25) and (27) through (32) with remaining
mismatches.
(34) Repeat steps (3) through (31). Note: Initial reference will be run with
DEVICE UNDER TEST port open.
(35) Record Measuring receiver indication in table F3.
Table F3. Return Loss Measurements (without N-Short)
Test
Mismatch
Mismatch
Mismatch
Mismatch
frequency
2334-001-1
2334-001-2
2334-001-3
2334-001-4