Alternate Calibration Process for Coaxial Circulators - Continued
Alternate Calibration Process for Power Splitters and Power Dividers (10 MHz to 18 GHz)

Calibration Procedure For Directional Couplers, Vswr Bridges, Coaxial Circulators, Power Splitters, And Power Dividers (10 Mhz To 40 Ghz) (General) Page Navigation
  44    45    46    47    48  49  50    51    52    53    54  
TB 9-5985-314-24
(c) OUTPUT PORT to 50 : termination.
(35) Measuring receiver indication will be 20 dB or greater.
b. Adjustment. No adjustments can be made.
2. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
G-4