Alternate Calibration Process for Coaxial Circulators
Final Procedure

Calibration Procedure For Directional Couplers, Vswr Bridges, Coaxial Circulators, Power Splitters, And Power Dividers (10 Mhz To 40 Ghz) (General) Page Navigation
  43    44    45    46    47  48  49    50    51    52    53  
img
. . . .
TB 9-5985-314-24
(a) OUTPUT PORT to POINT A.
(b) INPUT PORT to POINT B.
(c) SIDE PORT to 50 : termination.
(27) Measuring receiver indication will be 20 dB or greater.
(28) Repeat steps (17) through (24) but replace step (20) with the following:
Connect TI into figure 7 equipment setup as listed in (a) through (c) below:
(a) OUTPUT PORT to POINT A.
(b) SIDE PORT to POINT B.
(c) INPUT PORT to 50 : termination.
(29) Measuring receiver indication will be 0.5 dB or less.
(30) Repeat steps (17) through (24) but replace step (20) with the following:
Connect TI into figure 7 equipment setup as listed in (a) through (c) below:
(a) SIDE PORT to POINT A.
(b) OUTPUT PORT to POINT B.
(c) INPUT PORT to 50 : termination.
(31) Measuring receiver indication will be 20 dB or greater.
(32) Repeat steps (17) through (24) but replace step (20) with the following:
Connect TI into figure 7 equipment setup as listed in (a) through (c below:
(a) SIDE PORT to POINT A.
(b) INPUT PORT to POINT B.
(c) OUTPUT PORT to 50 : termination.
(33) Measuring receiver indication will be 0.5 dB or less.
(34) Repeat steps (17) through (24) but replace step (20) with the following:
Connect TI into figure 7 equipment setup as listed in (a) through (c) below:
(a) INPUT PORT to POINT A.
(b) SIDE PORT to POINT B.
G-3