Table F2. Return Loss Measurements (with N-Short)
APPENDIX G. ALTERNATE CALIBRATION PROCESS FOR COAXIAL CIRCULATORS (10MHZ TO 18GHZ)

Calibration Procedure For Directional Couplers, Vswr Bridges, Coaxial Circulators, Power Splitters, And Power Dividers (10 Mhz To 40 Ghz) (General) Page Navigation
  40    41    42    43    44  45  46    47    48    49    50  
img
. . .
TB 9-5985-314-24
(36) Add readings from table F2 and table F3 then divide by 2. Return loss in dB
indication will be within the limits specified in table 2 (for secondary reference as stated in
test report plus or minus specifications of SWR bridges).
2. Final Procedure
a. Deenergize and disconnect all equipment and reinstall protective cover on TI.
b. Annotate and affix DA label/form in accordance with TB 750-25.
F-5/(F-6 Blank)